Image Wisely, a joint initiative of ACR, RSNA, ASRT and AAPM,
provides information to the medical community to promote safety in medical imaging.
Patient injuries from fluoroscopic procedures should be reported in the same manner as other harmful events. Internal reporting structures vary by department or health care facility. Typically, reporting requirements are designed to meet certain standards of care that may be unique to that particular facility, but they are also designed to meet institutional external reporting requirements.
In some institutions the goal of collecting the data is to drive improvements; as such, those institutions may collect and report on data beyond what is required for reporting to outside entities. Health care providers should familiarize themselves with their institutional reporting structure and requirements, as well as those of external reports to state agencies.
External reporting can vary by state. The Joint Commission and the U.S Food and Drug Administration (FDA) also differ in their guidelines for reporting. The specific requirements for reporting fluoroscopic-related injuries will be discussed in this document.
In this document we use rules in the state of Massachusetts as an example. Although reporting requirements in other states may differ, familiarity with this state’s rules may help the reader navigate through their own state’s requirements [1].
The Massachusetts Department of Health and Human Services website provides information regarding X-ray imaging in “105 CMR 120.400: X-Rays in the Healing Arts” [2]. This document states that:
In our institution, we comply with these regulations by reporting weekly to the hospital radiation safety committee on all cases in which a patient receives a dose >2 Gy. We perform patient follow-up for all patients who receive a skin dose >5 Gy.
The Veteran’s Administration (VA) requires reporting on doses >3 Gy: “If 3 Gy skin dose to single field is exceeded, estimated skin doses and locations must be recorded in the patient’s medical record [3].”
In California, state law on CT and radiation therapy exams requires hospitals to report “Events” that fall into categories defined by the state (Section 115113). “CT of therapeutic exposure that results in unanticipated permanent functional damage to an organ or physiological system, hair loss, erythema, as determined by a qualified physician,” but fluoroscopy is not mentioned specifically [4].
The Board of Registration in Medicine in the State of Massachusetts under section 243 CMR 3:08 requires that “Major Incidents” be reported by the health care facility to Board of Registration in Medicine. There are several general categories that define a “Major Incident” [5].
Although not specifically mentioned by name, a fluoroscopic injury may fall under the following category: “All deaths or major or permanent impairments of bodily functions that are not ordinarily expected as a result of the patient’s condition on presentation.” Reporting of such events is required.
The Massachusetts Department of Public Health (DPH) requires that hospitals must report serious injury resulting from accidental or unknown causes and other incidents that seriously affect the health and safety of patients. In this context, “serious injury” means injury that is life threatening, results in death, or requires a patient to undergo significant additional diagnostic or treatment measures. “Accidents” include falls, burns, electrocutions and other misadventures not related to patient treatment.
If the specific fluoroscopic injury does not fall into the above two categories, it may fit into “Other serious incidents that seriously affect the health and safety of patients,” which can result in a serious injury not anticipated in the normal course of events [6].
The Joint Commission classifies radiation overdose as a Reviewable Sentinel Event [7,8]. In regard to fluoroscopy, the Joint Commission considers a Sentinel Event to be “prolonged fluoroscopy with cumulative dose >1500 rads to a single field” and defines cumulative dose as a dose given within a period of six months to a year. The Joint Commission defines a “single field” as the location on the skin through which fluoroscopic beam is directed (which can be from different projections), in an attempt to assess the maximum or peak skin dose the patient received. If this threshold is exceeded, the health care organization is required to conduct a root cause analysis and is encouraged to voluntarily report the event to The Joint Commission, even though the outcome (harm to the patient) has not become evident [9].
Health care providers are encouraged to file a voluntary report to the FDA through MedWatch if they suspect a problem with a fluoroscopic device or any event occurs that reasonably suggests a probability that a medical device has caused or contributed to the death, serious injury or serious illness of a patient. Such problems may include user error and device malfunction [10].